This meter is designed as wide frequency to supervise high frequency transmission of range from 50MHz to 3.5GHz.
|
This meter is designed as wide frequency to supervise high frequency transmission of range from 50MHz to 3.5GHz. It has the non-directional measurement feature and high sensitivity. It can be used to measure electromagnetic field intense of the transverse electromagnetic wave cell and Absorber rooms.
Specifications: Measurement method:Digital, triaxial measurement Directional characteristic:Isotropic, triaxial Display resolution:0.1mV/m, 0.1 μA/m, 0.1 μW/m², 0.001 μW/cm² Display type:LCD 4 digits Audible warning: buzzer Measurement units:mV/m, V/m , μA/m , mA/m , μW/m² , mW/m² , μW/cm² Measurement display value:Instantaneous measured value, maximum value or average value or maximum average value Measurement alarm function:Adjustable threshold with ON/OFF Measurement calibration factor CAL:Adjustable Manual data memory and read storage:70 sets (only recall by meter) Sensor type:Electric field (E) Frequency range:50MHz to 3.5 GHz Measurement range: Continuous(CW)signal(f>50MHz): 20mV/m to 108.0V/m; 53uA/m to 286.4mA/m; 1uW/m² to 30.93W/m²; 0uW/cm² to 3.093mW/cm² Dynamic range: about 65dB Absolute error(below 1V/m and 50MHz):± 1.0dB(for1.93GHz)
Frequency response: Taking into account the sensor CAL factor: ± 3.0dB (50MHz~800MHz) ; ±2.0dB(800MHz ~2.9GHz) ; ± 3.0dB (2.9GHz~ 3.5GHz) ;
Isotropy deviation: about ±1.0dB (f>50 MHz) Overload limit:4.2W/m2 (40 V/m) Temperature response(0 to 50 ℃) :± 0.2dB |